Potentiometric Glucose Detection by Paper-based Electrochemical Sensor on CMOS Chip

Katsuyuki Yamaoka, Jun Eguchi, Shigeyasu Uno

Abstract


This paper presents a low cost portable medical device for biochemical sensor using CMOS chip and paper-based fluidic channel. We measured a potential produced by enzyme activity of glucose between the working and reference electrode on CMOS chip. A liquid sample is transported by paper-based fluidic channel, which is made of chromatography paper and silicone resin, and consists of the area for filtering a sample (filter layer) and that for reacting enzyme (enyzme layer). The paper-based fluidic channel is used by combining CMOS chip, and the solution with glucose is dropped from top of the paper-based fluidic channel. The concentrations of glucose are detected by potentiometry (open circuit potential time). The experimental results show that the glucose concentration is measured by CMOS chip and paper-based fluidic channel.


Full Text:

PDF


DOI: http://doi.org/10.12928/telkomnika.v15i2.6129