Simultaneous Electrochemical Measurement using Paper Fluidic Channel on CMOS Chip
Jun Eguchi, Katsuyuki Yamaoka, Shigeyasu Uno
Abstract
This paper described the new system of biosensing using CMOS chip. The system was expected to be used in various circumstances because it was suitable for miniaturization compared to the conventional system. To conduct electrochemical measurements, the new system used paper fluidic channel set on the CMOS chip to transport solution to the on-chip electrodes. The materials of paper fluidic channel were only paper and silicone resin, and these were biocompatible. In experiment, we carried out simultaneous detection of glucose and ethanol in liquid sample solutions on the 5mm square CMOS chip and paper fluidic channel. Furthermore, this system can detect various target molecules in addition to glucose and ethanol, and increase number of simultaneous measurement by adding some more process to the paper and CMOS chip.
DOI:
http://doi.org/10.12928/telkomnika.v15i2.6131
Refbacks
There are currently no refbacks.
This work is licensed under a
Creative Commons Attribution-ShareAlike 4.0 International License .
TELKOMNIKA Telecommunication, Computing, Electronics and Control ISSN: 1693-6930, e-ISSN: 2302-9293Universitas Ahmad Dahlan , 4th Campus Jl. Ringroad Selatan, Kragilan, Tamanan, Banguntapan, Bantul, Yogyakarta, Indonesia 55191 Phone: +62 (274) 563515, 511830, 379418, 371120 Fax: +62 274 564604
<div class="statcounter"><a title="Web Analytics" href="http://statcounter.com/" target="_blank"><img class="statcounter" src="//c.statcounter.com/10241713/0/0b6069be/0/" alt="Web Analytics"></a></div> View TELKOMNIKA Stats